Product Description:
TC5H high temperature CO2 sensor is a new smart probe for measuring CO2 concentration based on dual-wavelength infrared technology
for demanding applications such as life science incubators.
The probe can be used for humid heat sterilization in the incubator, with dual output of analog and digital signals,
and real-time detection of changes in carbon dioxide concentration.
Features:
1. The measurement range is diverse and the accuracy is high
PPM level: 2000PPM, 5000PPM
VOL level: 10% VOL, 20% VOL, 100% VOL, etc
2. NDIR dual wavelength principle, with an expected lifespan of over 10 years
Infrared optical principle, probe has a service life of up to 10 years
Dual wavelength detector with excellent long-term stability
3. High temperature resistance, can be used for moist heat sterilization in incubators
Probe light path self heating (+2 ℃...+3 ° C) to prevent condensation in high humidity environments. The probe is resistant to high temperatures and can be used for moist heat sterilization in incubators
Description:
TC5H is a new intelligent probe based on dual channel infrared optical technology, designed for measuring carbon dioxide in harsh application environments such as life science incubators, which require stable, reliable, and accurate measurement performance. This probe has the following characteristics:
-Higher temperature tolerance, non working storage temperature can reach up to 95 ℃
-Wider measurement range, minimum 0... 2000ppm, maximum 0... 100% VO1
-Higher measurement resolution, with a resolution of up to 0.01% vol
-Higher measurement accuracy, with an error of less than+5% FS across the entire range
-Typical measurement range 0... 20% vol, performance optimization has been carried out for 5% CO2 measurement
-The signal adopts dual outputs of analog and digital, making it easier for users to use
-Probe light path self heating to prevent condensation in high humidity environments
-Long lifespan, with an expected lifespan of over 10 years
-Threaded casing and aviation plug interface, supporting multiple fixing methods, equipped with nuts/brackets
Parameters:
Working temperature: 0... 50 ℃
Working humidity: 0... 99% RH (no condensation)
Work pressure: 80... 120kpa
Storage temperature: -40... 95 ℃ (no condensation)
Preheating time: 30s; Suggest collecting data after 60 seconds of power on. Operating voltage: 5V DC
Working current:< 150mA
Range: 0... 10% VO1, 0 20%vol
Resolution: 0.01% vol
Analog signal output: 0.4... 2V corresponds to 0... 100% FS
Digital signal output: Modbus RTU
Accuracy:
When ≤ 50% FS:+ 3%FS;@ 5%CO2:+ 0.1%vol
>At 50% FS: ± 5% FS
Maintenance interval: every 12 months
Packing list: Sensor * 1, Instruction Manual * 1, Bracket * 1